Errors
Errors
Code Name
Meaning
Action
E1
Cartridge The cartridge is empty.
Change the cartridge immediately.
Empty
•
Press
the
Check
key twice to turn
off and confi rm the error.
• Disconnect or remove the infusion
set from your infusion site.
• Change the cartridge and prime the
infusion
set.
• Reconnect the infusion set (tubing)
at your infusion site.
•
Put your insulin pump into RUN
mode.
E2
Battery
The battery is depleted.
Replace the battery immediately.
Depleted
•
Press
the
Check
key twice to turn
off and confi rm the error.
• Replace the battery.
• Put you insulin pump into RUN mode.