49
Alert
Code Name
Meaning
Action
A4
Alarm
Goes off at the time you have
• Press the
Check
key twice to turn
Clock
programmed on the pump.
off and confi rm the alert.
• Put your insulin pump into RUN
mode if necessary.
A6
TBR
A temporary increase or
• Press the
Check
key twice to turn
Cancelled
decrease in the basal rate has
off and confi rm the alert.
been cancelled.
• Put your insulin pump into RUN
Note: Make sure that you
mode if necessary.
intended to cancel the TBR,
and if necessary, program a
new
TBR.
A7
TBR
A Temporary Basal Rate has
• Press the
Check
key twice to turn
Over
ended.
off and confi rm the alert.
Note: Your original basal rate
(100%) will resume automatically.
A8
Bolus
A bolus has been cancelled
• Press the
Check
key twice to turn
Cancelled
during the start delay or after
off and confi rm the alert.
delivery has begun.
• Put your insulin pump into RUN
Note: Make sure that you
mode if necessary.
intended to cancel the bolus,
and if necessary, program a new bolus.