48
Alerts
Alert
Code Name
Meaning
Action
A1
Cartridge The cartridge is almost empty
• Press the
Check
key twice to turn
Low
(it has only 20 units of U100
off and confi rm the alert.
insulin remaining).
• Change the cartridge before
it’s
empty.
A2
Battery
The battery is low.
• Press the
Check
key twice to turn
Low
off and confi rm the alert.
• Change the battery within the
next few hours.
• Put your insulin pump into RUN mode.
A3
Review
The battery has been removed
• Press the
Check
key twice to turn
Time and for more than 1 hour and the
off and confi rm the alert.
Date
time and date need to be
• Review the time and date and
re-entered.
make any necessary changes.