ABEM Terrameter LS
34
than one expansion unit is needed they must be of type ES10-64C (orange colour), the
older ES10-64 (grey colour) cannot link to other units.
The ES10-64 should be connected at the AUX connector via a multifunction cable
(ABEM part no 33 0020 11). The distance between the Terrameter LS and the first
ES10-64 is limited by the length of the multifunction cable (33 0020 11). In case it is
not long enough, an ”ES10-64 Communication Adapter” (33 0022 81) plus an “ES10-
64 Interlink Cable” is required.
Note!
Some ES10-64 units have a start-up problem. They
consume more power (12V DC) than expected during
start. This can be solved by using external power, or by
doing a minor hardware modification of the ES10-64
controller board. Please contact ABEM in case you
experience this problem.
Instrument software version 1.5.1 or higher is required to use an external relay switch.
The software will attempt to connect with the ES10-64 at measuring start, provided
the selected spread and protocol file demands an external relay switch. The spread
files must hold information about the external switching unit, as described in
“Appendix C. Spread and Measuring Sequence Files”.
4.6
The Power Source
The power supply
View
(Figure 37) shows the status for the power supply and internal
temperature of the instrument. The actual values are shown and they are
complemented with minimum and maximum values within square brackets.
Figure 37. Status view for power supply and temperature