Q 14: What is the “peak” format?
Answer One-point data is composed of both maximum and minimum values.
Details
Maximum and minimum points are detected at the maximum rate of amp. This controls the
number of data and allows completely catching changes in a signal. The following data is
peak format. Extension of .FPP and .ID signifies the peak format.
*The file, when a peak file is saved once again, with .DRT extension is also peak format.
Real-time record to recording paper also uses peak format. No data is lost by changing the
paper feeding speed.
Q 15: How to set the recorder for unmanned operation?
Answer The Timer function of System allows this measurement.
Details
The following is setting examples.
Example 1: Recording from 12:00, Jan 7 to 10:00, Jan 8 in 2005
Example 2: Recording from 9:00 to 10:00 every day
Example 3: Recording for 10 minutes every three hours for a month from now on.
Case of Example 1
Start Timer
Start date 2005/12/05/ 12:00
End date 2005/12/06/ 10:00
(A recording interval is not used)
Case of Example 2
Start Timer
Start date **/**/**** 9:00
Start date **/**/**** 10:00
(A recording interval is not used)
Case of Example 3
Start Timer
Start date 2005/01/05/12:00 (specify the time after the current time)
End date 2005/02/06/12:00 (specify the time at the finish time of a measurement)
Recording interval
Recording starting interval: 2 hours.
Recording operation interval: 10 minutes
The Start button is automatically pushed at the time specified in the start timer, and then a
recording operation is initiated. After starting, this operation works according to the specified
interval and ending time
Comment:
In combination with the Trigger button function, Trigger detection can be waited during
specified time.
- If the Trigger mode is turned off in Memory Recorder mode, the memory auto-save takes
effect during Stop. This allows automatically saving memory data based on the Stop time.
Caution:
Recording is not initiated with only Recording Interval Function ticked. To use interval,
make sure to configure the Start Timer and tick.
Summary of Contents for Omniace III RA2300A
Page 1: ...RA2300A OmniaceⅢ INSTRUCTION MANUAL 1AV7001754 R01 ...
Page 2: ...OmniaceⅢ RA2300A Instruction Manual ...
Page 16: ......
Page 17: ...1 RA2300A Overview ...
Page 21: ...2 Name and Function of Each Block ...
Page 28: ......
Page 29: ...3 Pre measurement Procedures ...
Page 40: ......
Page 41: ...4 Operation Flow Flow of Measurement Basic Settings and Operations ...
Page 48: ......
Page 49: ...5 Input Signal Monitor Observing Input Signals ...
Page 54: ......
Page 55: ...6 Auto Setup Automatically Setting Recording Conditions ...
Page 58: ......
Page 59: ...7 Amp Units ...
Page 62: ......
Page 63: ...8 Pen Recorder Recording Low speed Signal for Long Time ...
Page 68: ......
Page 69: ...9 Memory Recorder Recording High speed Signals ...
Page 76: ......
Page 77: ...10 HD Recorder Recording Data in Internal HDD ...
Page 85: ...11 Multi Recorder Separately Executing Waveform Printing Memory Recording and HD Recording ...
Page 91: ...12 X Y Recorder ...
Page 97: ...13 Trigger Settings Capturing Target Data to be Recorded ...
Page 108: ......
Page 109: ...14 Replay Display Displaying Recording Data ...
Page 121: ...15 Display and Printing Settings for Monitor Display and Printing ...
Page 124: ......
Page 125: ...16 System Setup Other Functions ...
Page 144: ......
Page 145: ...17 How to Use Optional Units ...
Page 156: ......
Page 157: ...18 Maintenance and Cleaning ...
Page 160: ......
Page 161: ...19 Troubleshooting ...
Page 168: ......
Page 169: ...20 Specifications ...
Page 188: ...20 Specifications RA2300A Recorder 7001754 R01 20 20 5 Event Box ...
Page 189: ...21 Cables Probes Spare Parts List ...