Entering the reference and tolerance values
10.
When is displayed, press the PRINT key.
will be displayed.
11.
When is displayed, press the PRINT key to
check the currently set value (all blinking).
Press the RE-ZERO key to enter the load input mode.
12.
Press the RE-ZERO key,
0.00
g
is displayed. Place a
sample of the weight of the upper limit on the balance and
press the PRINT key. (Register the upper limit value.)
13.
When finished,
Cp lo
is displayed. (Replace a sample of the
weight of the upper limit from the balance.)
14. When
Cp lo
is displayed, press the PRINT key to check
the currently set value (all blinking). Press the RE-ZERO key to
enter the load input mode.
15.
Press the RE-ZERO key,
0.00
g
is displayed.
16. Place a sample of the weight of the lower limit on the balance
and press the PRINT key. (Register the lower limit value.)
17.
When finished,
Cp Hi
is displayed. (Replace a sample of the
weight of the lower limit from the balance.)
Press the CAL key twice to return to the weighing display.
Place a sample of
the weight of the upper limit
Replace a sample
x 2
Place a sample of
the weight of the lower limit