![A&D GF-10202M Instruction Manual Download Page 28](http://html.mh-extra.com/html/aandd/gf-10202m/gf-10202m_instruction-manual_446426028.webp)
28
6-3
Percent Mode (
%
)
The percent mode displays the weighting value in percentage compared to a 100%
reference mass and is used for target weighing or checking the sample variance.
Selecting The Percent Mode
1. Press the MODE key to select the unit % (Percent mode).
Storing The 100% Reference Mass
2. Press the SAMPLE key to enter the 100% reference
mass storing mode.
Even in the storing mode, pressing the MODE key will
switch to the next mode.
3. Place a container on the weighing pan, if necessary. Press the RE-ZERO
key to cancel the weight (tare). The balance displays
100 0
%
.
4. Place the sample to be set as the 100% reference mass on the pan or in
the container.
5. Press the PRINT key to store the reference mass.
The balance displays
100.00
%. (The decimal point
position depends on the reference value. The reference
mass stored, even if the AC adapter is removed, is
maintained in non-volatile memory.)
Note
□
If the balance judges that the mass of the sample is too light to be
used as a reference, it displays
lo
.
□
The displayed percentage is based on the 100% reference mass.
Model
100% mass
Decimal point position
Minimum display 0.01g 1.00g ~ 9.99g
1%
10.00g ~ 99.99g
0.1%
100.00g ~
0.01%
Minimum display 0.1g
10.0g ~ 99.9g
1%
100.0g ~ 999.9g
0.1%
1000.0g ~
0.01%
□
Registered values are stored even when the power is turned off.
6. Remove the sample
Reading the percentage
7. Please a sample to be compared to the reference mass on the pan.
The displayed percentage is based on the 100% reference mass.
place
Display % of weighing object
Weighing pan
Container
Sample
(100%)