Display
Error code
Description
Full memory
The stored calibration history has reached 50 instances. If
more is stored, the old history will be deleted. Refer to
"13.Data Memory"
EC,E00
Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC,E01
Undefined command error
An undefined command was received.
Confirm the command.
EC,E02
Not ready
A received command can not be processed.
Example:
The balance received a "Q" command, but not in the weighing
mode.
The balance received a "Q" command while processing a
RE-ZERO command. Adjust the delay time to transmit a
command.
EC,E03
Timeout error
If the timeout parameter is set to "
t-Up1
", the balance did
not receive the next character of a command within the time
limit of one second. Confirm the communication.
EC,E04
Excess characters error
The balance received excessive characters in a command.
Confirm the command.
EC,E06
Format error
A command includes incorrect data.
Example:
The data is numerically incorrect.
Confirm the command.
EC,E07
Parameter setting error
The received data exceeds the range that the balance can
accept. Confirm the parameter range of the command.
(
)
Blink
Alternate