107
3. Press the PRINT key to store.
If the flow rate unit is g / *, the display returns to weighing display or flow display.
If the flow rate unit is mL / *, the display transitions to density setting.
Please refer to "How to set the density".
* U
nit of time setting (second(s),minute(m) or hour(h)) is entered in
"
* " of "g/*" and "mL/*".
How to set the density
When the setting value of function setting
Frd Unit
is 3, 4, 5,
after setting the calculation time
or calculation precision, go to density
setting display.
Density can be changed by following key operation.
The setting range is 0.0001g/cm
3
to 9.9999g/cm
3
.
RE-ZERO (+) key ꞏꞏꞏꞏꞏ Change the number of the blinking digit
MODE (-) key ꞏꞏꞏꞏꞏꞏꞏꞏꞏ Change the number of the blinking digit
SAMPLE key ꞏꞏꞏꞏꞏꞏꞏ Move the blinking digit
PRINT key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Store the set value and return to weighing display.
CAL key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Return to weighing display without storing the set
value.
Method of reading density number
When flow unit is mL/*, up to 10 densities can be registered.
To register a new density, read the unconfigured density number
and then register according to the procedure of the setting method
of calculation time or calculation precision.
Continuing to hold down the PRINT key (for 2 seconds) in
weighing display displays .
Blinking is the current density number and is the set
density value.
The density number can be changed by following key operation
Note
f
** : The selected density number is entered.
d
**** : The set density number is entered.
The setting range is F01
~
F10.
RE-ZERO (+) key ꞏꞏꞏꞏ
Change density number.
MODE (-) key ꞏꞏꞏꞏꞏꞏꞏꞏꞏ
Change density number.
PRINT key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
Read the density of the s
elected density
number and return to weighing display.
CAL key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Return to the weighing display or flow display without
reading the density of the selected density number.
Change display
After returning to the weighing value display after setting to flow mode,
the unit is "g" with the or indicator on.
Use the MODE key to switch between flow rate display and "g"
display.
By switching, the total amount and flow rate can be checked.
.
.
Weighing display
From the Flow calculation time
or Flow calculation precision
Total amount
display
Flow rate
Press and hold
(for 2 seconds)