![A&D GF-10202M Instruction Manual Download Page 37](http://html.mh-extra.com/html/aandd/gf-10202m/gf-10202m_instruction-manual_446426037.webp)
37
Model
Usable calibration weight
Factory
setting
Adjustable
range
GX-8202M,
GF-8202M
2kg to 8kg (1kg interval)
5kg
-99.99g to
+99.99g
GX-8202MD,
GF-8202MD
2kg to 8kg (1kg interval)
5kg
GX-10202M,
GF-10202M
2kg to 10kg (1kg interval)
10kg
GX-12001M,
GF-12001M
5kg,
10kg
10kg
-999.9g to
+999.9g
GX-22001M,
GF-22001M
5kg,
10kg, 20kg
20kg
GX-32001M,
GF-32001M
5kg,
10kg, 20kg, 30kg
20kg
GX-32001MD, GF-32001MD 5kg,
10kg, 20kg, 30kg
20kg
Display
This indicator means "In process of measuring calibration data".
Do not allow vibration or drafts to affect the balance while the indicator is
displayed.
9-1
Automatic Self Calibration For The GX-M Series
This function automatically calibrates the balance using the internal weight due to a temperature
change in the ambient environment, time set or interval time. It can function even when the balance’s
display is in off. If GLP output in function table is selected, the balance outputs the sensitivity
adjustment report or store the data in memory after calibration.
In the auto calibration mode, either the temperature change (
CFnc 0
), the setting time (
CFnc 1
), or the
interval time (
CFnc 3
) can be set with the function setting
CFnc
.
For the setting time, the three function setting of , and can be set.
Interval time can be set from 0.5h to 24h with function setting .
Caution
If something is on the weighing pan, the balance judges that it is in use and does not perform automatic
self calibration.
The criteria that the balance judges is in use are as follows.
0.01g models
0.1g models
More than 20g
More than 200g
To maintain the calibrated state, keep the weighing pan clear while not in use.
Indicates that the balance detects a change in ambient temperature and
automatic self calibration will start. If the balance is not used for a few
minutes with this indicator ( ) blinking, the balance performs automatic self
calibration. The blinking duration depends on the environment.
Indicates that the balance is measuring calibration data. Do not allow
vibration or drafts to affect the balance while this indicator is displayed. After
calibration, the balance returns to indicate the previous display.
Note
The balance can be used while the indicator blinks. But, it is recommended that to
maintain the accuracy, stop using the balance and confirm that there is nothing on the pan
and allow the balance to perform self calibration.
Depending on the setting of "10.Function Switch And Initialization" , "change prohibited" or
"changeable (usable) " can be selected.
Summary of Contents for GF-10202M
Page 141: ...141 29 External Dimention Unit mm ...
Page 145: ...145 MEMO ...
Page 146: ...146 MEMO ...