Entering and outputting the minimum weighing value
17-2-1 Setting from function setting
【
Inputting the setting value directly
】
Continue from “Step 6” in “17-1 Comparing the minimum weighing value”
7. Press the PRINT key while
MW
is displayed.
8. At
KEY in
display
When setting minimum weighing value
Press the PRINT key again and proceed to Step 9.
When not setting minimum weighing value
Press the CAL key twice to proceed to weighing
mode without entering minimum weighing value.
9. Set the minimum weighing value.
Use the following keys to change the minimum weighing value.
RE-ZERO (
+
) key
.....
Increase a numerical value of the digit being blinked
MODE (
-
) key
........ .....
Decrease a numerical value of the digit being blinked
SAMPLE key
......
Move the digit being blinked
PRINT key
.............
Store a parameter and proceed to next stage.
* In this time, when
MW-CP
is set to
0,
the balance automatically set
it to
1
(except near zero) and enables comparing function of the
minimum weighing value.
CAL key
..................
Proceed to next stage without storing the setting value.
10.
Press the
CAL key to return to weighing mode.
MW
MW
End
Mi
N
0.00
Un it
KEY in