8-14
Chapter 8: Maintenance
Align the Confocal KLH-1 Lamphouse
To align the confocal KLH-1 lamphouse (Figure 8-10):
Figure 8-10. KLH-1 Lamphouse Adjustment Knobs
1
2
1
Vertical lamp
adjustment knob
2
Horizontal lamp
adjustment knob
e
Lamp condenser
adjustment knob
3
1.
Use the horizontal lamp knob on the lamphouse to bring the scan line just
into view at the bottom of the screen.
2.
Use the vertical lamp knob on the lamphouse to flatten the scan line.
3.
Move back and forth between the horizontal and vertical lamp knobs to
achieve a flat scan line (Figure 8-11).
Figure 8-11. Flat Scan Line
1
1
scan line
Содержание KMS-310
Страница 10: ...Contents viii...
Страница 11: ...About This Manual Intended Audience Manual Organization Conventions Contacting Technical Support 1...
Страница 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Страница 51: ...User Interface Overview Software Controls Software Organization 4...
Страница 53: ...KMS 310 400 Supervisor s Manual 4 3 Figure 4 1 Key Software Elements 1 2 3 4 5 6 7 1 2 3 4 5 6 7...
Страница 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Страница 219: ...Operation Overview Access and Exit the Supervisor s Menu Set up a Program Measure Manage Measurement Data 7...
Страница 266: ...7 48 Chapter 7 Operation...
Страница 267: ...Maintenance Overview Lamp Replacement and Alignment Stage Maintenance Electronics Cabinet Maintenance System Cleaning 8...
Страница 292: ...8 26 Chapter 8 Maintenance...
Страница 293: ...Error Messages System Error Messages Script Error Messages 9...
Страница 297: ...Glossary...
Страница 304: ...I 4 Index...