![ZYGO KMS-310 Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/zygo/kms-310/kms-310_supervisor-manual_943104038.webp)
3-6
Chapter 3: System Overview
are not switched for system components. All of these outlets are deactivated
whenever the emergency off (EMO) button is pushed.
The Piezo control assembly is used to fine focus a substrate to an objective by
moving the substrate toward or away from the objective in increments of 0.01-
0.02 microns. Moving the thumbwheel on the keypad up will move the substrate
closer to the objective, and moving it down will move the substrate away from
the objective.
The Piezo control assembly should be kept at the center of its range ±15 microns
for transmitted light systems (±10 microns for reflected or confocal light
systems) until you are approximately in focus with the column (coarse) focus.
Once you have completed coarse focus, you can move the Piezo to find the
desired feature and perform autofocus.
Power Control Panel
The power control panel is located in the front of the electronics cabinet
(Figure 3-4). Functions for each of the control panel components are listed in
Table 3-1.
Figure 3-4. Power Control Panel
1234
Table 3-1. Power Control Panel Component Functions
Figure Callout
Component
Function
1
Power On
Switch
This switch is the master control switch for the
entire system. It controls power to all subsystems
within the KMS-310/400 system. (The lamp power
supply must be turned on separately before
applying power to the system.)
2
Reset switch
This switch resets the entire system and reboots
the computer. When this switch is activated, you
will need to reinitialize the stage and program
operations.
3
Power Off
switch
When this switch is pressed, all power to the
system is removed. (Power to the lamps is still on.)
4
Keyboard
connector
This connector allows you to connect a keyboard
to the system, which you can use in place of or in
addition to the standard keypad.
1
power on switch
2
reset switch
3
power off switch
4
keyboard connector
Содержание KMS-310
Страница 10: ...Contents viii...
Страница 11: ...About This Manual Intended Audience Manual Organization Conventions Contacting Technical Support 1...
Страница 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Страница 51: ...User Interface Overview Software Controls Software Organization 4...
Страница 53: ...KMS 310 400 Supervisor s Manual 4 3 Figure 4 1 Key Software Elements 1 2 3 4 5 6 7 1 2 3 4 5 6 7...
Страница 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Страница 219: ...Operation Overview Access and Exit the Supervisor s Menu Set up a Program Measure Manage Measurement Data 7...
Страница 266: ...7 48 Chapter 7 Operation...
Страница 267: ...Maintenance Overview Lamp Replacement and Alignment Stage Maintenance Electronics Cabinet Maintenance System Cleaning 8...
Страница 292: ...8 26 Chapter 8 Maintenance...
Страница 293: ...Error Messages System Error Messages Script Error Messages 9...
Страница 297: ...Glossary...
Страница 304: ...I 4 Index...