5-20
Chapter 5: System Setup
3.
To set up the measurement window with the NIST SRM 475 line width
standard mask:
a.
Center the fiducial line of Target 5 on the Y-axis.
b.
Adjust the measurement windows so that they are no higher or lower
than the fiducial mark on the right side of the pitch
(Figures 5-14 and 5-15).
c. Press
the
Recall
button to save this window in Recall 1.
Figure 5-14. NIST SRM-475 Standard Target 5D
Pitch
Figure 5-15. NIST SRM-475 Standard Target 5D
Line
4.
To enter the calibration standards from the Calibration Control submenu:
a.
From the Supervisor Master Menu screen, highlight the
Calibration
icon and press [SELECT].
b.
Input the pitch and line values into the corresponding
Standard
Horizontal
and
Standard Vertical
fields.
Содержание KMS-310
Страница 10: ...Contents viii...
Страница 11: ...About This Manual Intended Audience Manual Organization Conventions Contacting Technical Support 1...
Страница 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Страница 51: ...User Interface Overview Software Controls Software Organization 4...
Страница 53: ...KMS 310 400 Supervisor s Manual 4 3 Figure 4 1 Key Software Elements 1 2 3 4 5 6 7 1 2 3 4 5 6 7...
Страница 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Страница 219: ...Operation Overview Access and Exit the Supervisor s Menu Set up a Program Measure Manage Measurement Data 7...
Страница 266: ...7 48 Chapter 7 Operation...
Страница 267: ...Maintenance Overview Lamp Replacement and Alignment Stage Maintenance Electronics Cabinet Maintenance System Cleaning 8...
Страница 292: ...8 26 Chapter 8 Maintenance...
Страница 293: ...Error Messages System Error Messages Script Error Messages 9...
Страница 297: ...Glossary...
Страница 304: ...I 4 Index...