9-2
Chapter 9: Error Messages
System Error Messages
This section provides an explanation for common KMS-310/400 system error
messages.
Table 9-1 lists common system error messages. For each error message, this table
provides a definition and suggested course of action to correct the problem.
Table 9-1. System Error Messages
Error Message
Meaning
Corrective Action
Z-column
error #6
Tried to drive Z-Column below
mechanical limit switch
1.
Remove the mask.
2.
Power down the system.
3.
Restart the system.
Z-column at
upper limit
Tried to drive the objectives up
and too far away from the
sample
Select the
Focus
icon and drive
the column back down until the
message goes away.
Z-column at
lower limit
Tried to drive the objectives
down too close to the sample
Select the
Focus
icon and drive
the column away from the
sample until the message goes
away.
Z-column
software limit
Z software limit has been set
during objective setup. The
operator has attempted to drive
towards the sample and past the
software limit.
1.
Drive the column back
away from the sample.
2.
If desired, reset to a larger
software limit value by
performing the objective
setup procedure.
Illegal turret
move
Virtual objective was set in the
System.cfg file and the objective
set up was not carried out
properly.
Reset the objective setup. Take
care to correctly set the
relationship between the
reference objective and RIC
objective plane(s).
Edges not
found
Edge selection for given feature
is incorrect.
Operator has picked the wrong
edge count for a given feature.
Select the
Edges
icon, re-select
the correct edges, and scan
direction for the feature.
0.000 for
measurement
value
Measurement was taken on one
edge only,
or
measurement was out of range
for the current multi-point (sub-
M) file.
1.
Select the
Edges
icon, re-
select the correct edges, and
scan direction for the
feature.
2.
Check the expected feature
CD value against the multi-
point calibration table
limits.
Содержание KMS-310
Страница 10: ...Contents viii...
Страница 11: ...About This Manual Intended Audience Manual Organization Conventions Contacting Technical Support 1...
Страница 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Страница 51: ...User Interface Overview Software Controls Software Organization 4...
Страница 53: ...KMS 310 400 Supervisor s Manual 4 3 Figure 4 1 Key Software Elements 1 2 3 4 5 6 7 1 2 3 4 5 6 7...
Страница 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Страница 219: ...Operation Overview Access and Exit the Supervisor s Menu Set up a Program Measure Manage Measurement Data 7...
Страница 266: ...7 48 Chapter 7 Operation...
Страница 267: ...Maintenance Overview Lamp Replacement and Alignment Stage Maintenance Electronics Cabinet Maintenance System Cleaning 8...
Страница 292: ...8 26 Chapter 8 Maintenance...
Страница 293: ...Error Messages System Error Messages Script Error Messages 9...
Страница 297: ...Glossary...
Страница 304: ...I 4 Index...