6.
Close down the aperture so it is completely visible in the ocular eyepieces at
10X (Figure 5-2).
Figure 5-2. Aperture in Ocular View and Focus
Aperature partially closed while
viewing in the oculars. Image of
aperture is out of focus.
Aperture completely closed while
viewing in the oculars. Image of
aperture is in focus.
7.
While viewing the aperture in the oculars, turn the rear knob on the
transmitted light assembly until the image is in focus (Figure 5-2).
The best focus is found when neither a blue or red flare is seen around the
aperture when seen through the oculars.
Содержание KMS-310
Страница 10: ...Contents viii...
Страница 11: ...About This Manual Intended Audience Manual Organization Conventions Contacting Technical Support 1...
Страница 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Страница 51: ...User Interface Overview Software Controls Software Organization 4...
Страница 53: ...KMS 310 400 Supervisor s Manual 4 3 Figure 4 1 Key Software Elements 1 2 3 4 5 6 7 1 2 3 4 5 6 7...
Страница 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Страница 219: ...Operation Overview Access and Exit the Supervisor s Menu Set up a Program Measure Manage Measurement Data 7...
Страница 266: ...7 48 Chapter 7 Operation...
Страница 267: ...Maintenance Overview Lamp Replacement and Alignment Stage Maintenance Electronics Cabinet Maintenance System Cleaning 8...
Страница 292: ...8 26 Chapter 8 Maintenance...
Страница 293: ...Error Messages System Error Messages Script Error Messages 9...
Страница 297: ...Glossary...
Страница 304: ...I 4 Index...