KMS-30/400 Supervisor's Manual
5-19
2.
To set up the measurement window with the NIST SRM 473 line width
standard mask:
a.
Center the fiducial line of target 5D on the Y-axis.
b.
Adjust the measurement windows so that they are no higher or lower
than the fiducial mark on the right side of the pitch
(Figures 5-12 and 5-13).
c.
Press the
Recall
button to save this window in Recall 1.
Figure 5-12. NIST SRM-473 Standard Target 5D
Pitch
Figure 5-13. NIST SRM-473 Standard Target 5D
Line