KMS-310/400 Supervisor’s Manual
2-10
Objective Anti-Crash Protection
To avoid crashing an objective on the KMS-310/400 system, perform the
following procedure during initial system setup or after performing preventive
maintenance:
1.
With the main menu displayed, select
.
2.
Use the 100X or 50X magnification objective to focus on the highest
(upmost) feature of the image.
3.
With the image in focus, note the Z-axis scale reading (in microns).
4.
Use the
Ã
key to drive the objective 100 microns closer to the image.
For example, the Z-axis scale readings should increase by 100 microns.
5.
Use the thumbwheel (on the side of the column) to move the Z-axis limit
switch until the red LED on the side of the column illuminates.
The Z-axis lower limit switch is now set.
6.
Refocus the 10X objective.
Содержание KMS-310
Страница 10: ...Contents viii...
Страница 11: ...About This Manual Intended Audience Manual Organization Conventions Contacting Technical Support 1...
Страница 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Страница 51: ...User Interface Overview Software Controls Software Organization 4...
Страница 53: ...KMS 310 400 Supervisor s Manual 4 3 Figure 4 1 Key Software Elements 1 2 3 4 5 6 7 1 2 3 4 5 6 7...
Страница 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Страница 219: ...Operation Overview Access and Exit the Supervisor s Menu Set up a Program Measure Manage Measurement Data 7...
Страница 266: ...7 48 Chapter 7 Operation...
Страница 267: ...Maintenance Overview Lamp Replacement and Alignment Stage Maintenance Electronics Cabinet Maintenance System Cleaning 8...
Страница 292: ...8 26 Chapter 8 Maintenance...
Страница 293: ...Error Messages System Error Messages Script Error Messages 9...
Страница 297: ...Glossary...
Страница 304: ...I 4 Index...