5-8
Chapter 5: System Setup
Set Up Transmitted Illumination
This section provides instruction for setting up the transmitted illumination
controls in order to maximize the illumination signal while minimizing noise.
The basic steps required to set up transmitted illumination include:
setting illumination power (percent illumination)
adjusting camera gain (KMS-400RT system only)
configuring the electronics (frame grabber) controller for gain, offset and
analog gain
To set up transmitted illumination:
1.
Place an artifact on the stage and focus at 10X.
2.
Position a feature into the center of the measurement screen.
3.
From the Measure Screen, select the
Window
icon and adjust the
measurement window so that it falls across the feature on the screen.
4.
Highlight the
icon and press [MENU].
5.
Set the following values:
Gain = 31
Offset = 31
Analog = 1
6.
Use the arrows on either side of the number read-out bar to adjust the
percent illumination up or down so that the upper slide bar floats just below
the point it turns red.
When the percent illumination is correctly adjusted, the bar is white and will
move up and down slightly (Figure 5-5).
L
Note: You may need to slightly adjust the gain in order to correctly position the bar for
percent illumination.
Figure 5-5. Percent Illumination Read-out Bar
Before
After
Содержание KMS-310
Страница 10: ...Contents viii...
Страница 11: ...About This Manual Intended Audience Manual Organization Conventions Contacting Technical Support 1...
Страница 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Страница 51: ...User Interface Overview Software Controls Software Organization 4...
Страница 53: ...KMS 310 400 Supervisor s Manual 4 3 Figure 4 1 Key Software Elements 1 2 3 4 5 6 7 1 2 3 4 5 6 7...
Страница 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Страница 219: ...Operation Overview Access and Exit the Supervisor s Menu Set up a Program Measure Manage Measurement Data 7...
Страница 266: ...7 48 Chapter 7 Operation...
Страница 267: ...Maintenance Overview Lamp Replacement and Alignment Stage Maintenance Electronics Cabinet Maintenance System Cleaning 8...
Страница 292: ...8 26 Chapter 8 Maintenance...
Страница 293: ...Error Messages System Error Messages Script Error Messages 9...
Страница 297: ...Glossary...
Страница 304: ...I 4 Index...