3
NOTE:
The system has a built-in screensaver to prevent damage to the display. It is not recommended to continually turn the
unit on and off. To dispose of this product properly, please contact your local service department.
2.2
Electrical safety
Type of device protection against electric shock
CLASS I DEVICE
, where protection against electric shock is based not solely on BASE ISOLATION; the device also has
protective earthing. This additional safety precaution prevents exposed metal parts from becoming LIVE should there be
an insulation failure.
Degree of protection against electric shock
Type BF applied part
(for probes marked with the BF symbol)
TYPE BF APPLIED PART, providing a specified degree of protection against electric shock, in particular with regard to
the permitted leakage current.
BF: Insulation against earth; max. patient leakage current: under normal conditions < 100 μA,
in the event of the first failure < 500 μA
Degree of protection against harmful ingress of water
The parts of the probe that are likely to come in contact with the user or the patient meet the requirements for protection
against dripping water (IPX1). The parts of the probe that are intended for immersion in the course of normal use meet
the requirements for protection against temporary submersion (IPX7). The IP classification of the device is that for ordi-
nary devices (no protection, IPX0).
Safety level for use in oxygen-enriched environments, inflammable anaesthetics or other inflammable or com-
bustion-promoting substances
The device is not suitable for use in oxygen-enriched environments, inflammable anaesthetics or other inflammable or
combustion-promoting substances.