115
care when handling and protect the device from damage when not in use. NEVER use a damaged or DEFECTIVE
probe. Failure to follow these precautions can result in serious injury and equipment damage
Risk of electric shock:
The probe is operated using electrical energy that can injure the patient or user if live internal parts come into contact
with the conductive solution:
DO NOT immerse the probe in any liquid beyond the level indicated by the immersion depth marking. Never immerse
the probe connector in any liquid.
Prior to each use, visually inspect the probe lens and check the casing area for cracks, cuts, tears, and other signs of
physical damage. DO NOT use ANY probe that appears to be damaged until you have verified that it can be used
safely. Every time you clean the probe, you must perform a more thorough inspection, including the cable, strain relief
and connector.
Before inserting the connector into the probe port, inspect the probe connector pins. If a pin is bent, DO NOT use the
probe until it has been inspected and repaired/replaced by a ZIMMER MEDIZINSYSTEME representative.
The ZIMMER MEDIZINSYSTEME service team or qualified clinicians should routinely perform electrical leakage tests.
Mechanical risks:
A defective probe or excessive force can cause injury to the patient or damage to the probe:
Observe the depth markings and do not apply excessive force when inserting or guiding the endocavitary probe.
Inspect probes for sharp edges or rough surfaces that may injure sensitive tissue.
Do not apply excessive force to the probe connector when inserting into the probe port. This can bend the probe con-
nector pins.