Chapter 2 System safety
2.1
Safety overview
This section deals with measures for ensuring the safety of users and patients. In order to ensure the safety of users and
patients, please read the corresponding information in this chapter before putting the system into operation.
Failure to
observe the warnings or breach of the relevant regulations can lead to injuries, or even the loss of life, to users
or patients.
Users must observe the following precautions:
This system is of Type BF and complies with the applicable EN and IEC standards.
This system may not be modified in any way. Necessary changes to the device may only be carried out by the manufac-
turer or by its appointed representatives.
This system was fully calibrated at the factory. Do not adjust any fixed adjustable parts.
In the event of a malfunction, turn off the system immediately and inform the manufacturer or its designated agent.
The system's power cable should only be connected to an earthed protective power socket. Do not under any circum-
stances remove the earthing cable.
Only connect this system to electronic or mechanical devices that comply with the EN60601-1 standard. Check the
leakage current and other safety parameters of the entire system to avoid possible damage and risks caused by leak-
age currents and their overburdening.
The system does not have any special protective measures for the event that it is combined with high-frequency devices.
The operator should use extreme caution with these types of devices, as a matter of principle.
The system may only be installed by persons authorised by the manufacturer. Do not attempt to install the system your-
self.
Maintenance be only be performed by an authorised service technician.
The system may be used only by a qualified operator, or under qualified supervision.
Do not use this system in the vicinity of inflammable substances, otherwise an explosion may occur. Do not continu-
ously scan the same part of a patient's body or expose the patient to extensive scanning. This could harm the patient.
When using the system for ultrasound testing, use only appropriate ultrasound gel that complies with the system specifi-
cations.
Do not unplug the probe from the system when it is in active operation. Always go to the examination screen when
you need to remove the probe.
To prevent injury to the arms or the neck region, the operator should not stay in the same position for too long during the
examination.
Do not place any liquids on top of the device.