114
WARNING:
The probes are designed to be used with this ultrasound system only. Use of these probes on any other
system or use of any non-appropriate probe may cause an electric shock or damage to the sys-
tem/transducer.
9.2.2
Cleaning and disinfection
Disconnect the probe from the ultrasound console and remove all coupling gel from the probe by wiping it with a soft
cloth and rinsing with running water.
Wash the probe with mild soap in lukewarm water. If necessary, rub the probe with a soft sponge, gauze, or cloth to
remove any visible residue from the probe surface. Rinse the probe with enough clean, potable water to remove all
residues of disinfectant.
Use a soft cloth to clean the cable and the user section of the probe with a liquid cleaning disinfectant. Make sure that
the surface of the probe and the cable are sufficiently wet with the disinfectant cleaning agent.
Allow the probe to air dry completely.
Reconnect the probe to the ultrasound console and place the probe into its holder.
Immersion levels for the probe
CAUTION:
These transducers are not designed to withstand heat sterilisation procedures. Exposure to tempera-
tures in excess of 60ºC will cause permanent damage. The transducers are not designed to be totally
submerged in fluid, as permanent damage will result if the entire transducer is submerged.
Probe safety
Handling precautions
Ultrasound probes are highly sensitive medical instruments that can easily be damaged by improper handling. Use