126
Appendix A: Acoustic output report table
Acoustic power table
System: SonidoSmart
Transducer model: C3-A
Operating mode: B
Index sign
MI
TIS
TIB
TIC
Scan
No scan
No scan
Aaprt≤1c
m
2
Aaprt
>
1c
m
2
Global maximum
index value
0.5
0.1
#
Associated
p
ra
Mpa
0.92
acoustic
P
mW
2
#
parameters
Min. of [P
α
(Z
s
I
ta.α
(Z
s
)]
mW
Z
s
cm
Z
bp
cm
Z
b
cm
Z at max
pi.α
cm
3.2
d
eq
(Z
b
)
cm
f
awf
MHz
3.65
3.75
#
Attenuation
of A
aprt
X
cm
2.09
#
Y
cm
1.1
#
Other
t
d
μs
0.57
Information
prr
Hz
2293.6
p
r
at max.l
pi
MPa
1.23
d
eq
at max.l
pi
cm
I
pa.α
at max.MI
W/cm
2
35.21
Operating
Mode
B
B
#
control
Focus
cm
6
4
#
conditions
Output
%
100
100
#
NOTE 1: Data should only be entered in one of the columns relating to TIS.
NOTE 2: Information for the TRANSDUCER DEVICE that is not intended for trans-cranial or neonatal brain applications may not be pro-
vided in relation to TIC.
NOTE 3: If the requirements of 201.12.4.2a) are met, it is not necessary to enter data relating to TIS, TIB or TIC.
NOTE 4: If the requirement of 201.12.4.2b) is met, it is not necessary to enter data in the column relating to MI.