152
Voltage dips,
short-term inter-
ruptions and fluc-
tuations in the
supply voltage
in accordance with
IEC 61000-4-11
< 5 %
U
T
(> 95 % dip in
U
T
) for
½ period
40 %
U
T
(60 % dip in
U
T
) for 5
periods
70 %
U
T
(30 % dip in
U
T
) for
25 periods
< 5 %
U
T
(> 95% dip in U
T
) for 5
s
< 5 %
U
T
(> 95 % dip in
U
T
) for
½ period
40 %
U
T
(60 % dip in
U
T
) for 5
periods
70 %
U
T
(30 % dip in
U
T
) for
25 periods
< 5 %
U
T
(> 95% dip in U
T
) for
5 s
The supply voltage quality
must correspond to that of a
typical commercial or hospi-
tal environment. If the user of
the SONIDOSMART requires
continued functionality even
in the case of interruptions to
the power supply, it is rec-
ommended that the
SONIDOSMART be powered
from an uninterruptible power
supply or a battery.
Magnetic field at
the supply fre-
quency (50/60
Hz)
in accordance
with IEC 61000-
4-8
3 A/m
3 A/m
Magnetic fields at mains fre-
quency should have the typical
values found in a commercial
or hospital environment.
NOTE:
U
T
is the mains AC voltage before application of the test level..