Operation
43
Operating manual TEMPAR console V9.0
626140-9149-911
In the step "
Sensors
", the sensors of a console or measuring zone can be
selected, shifted, activated or deactivated.
A measuring zone is the area that is monitored using spatially arranged
and grouped sensors.
– Temperature sensors can be selected by clicking in the area "
Sensor
net configuration
" and then moved to another position in the
room. The change is displayed in parallel to this in the "Measuring
cube selection” area. The dimensions of the room can be entered in
metres (m) by clicking on "X”, "Y” or "Z”.
– All measuring zones and assigned sensors are displayed in the area
"
Measuring cube selection
". The currently selected measuring
zone is displayed in green. By selecting another measuring zone, its
configuration is displayed in the area "Sensor net configuration” and
changes can be made.
The grey highlighted area in "Measuring cube selection” marks the
local measuring room. This measuring cube is monitored from the
console. All sensors outside the local cube are recorded, but not
however monitored. They are used to define measuring zones in a
cockpit. By selecting a cube (which is not a local measuring cube) and
then clicking on the button "Define as local cube”, this cube becomes
the local measuring cube.
– In the "
Sensors
" area, sensors from the Hardware sensors area can
be assigned a sensor name (e.g. T1). The measured values recorded by
the sensor are then saved and displayed under this name. Such an
assignment should only take place in exceptional cases, e.g. during a
recalibration.
– Each sensor is represented as a tile and can be deleted via the top
right minus sign.
– Sensors can be selected or deselected in the column "
Gradient
calculation
". Deselected sensors are ignored in the calculation of key
quantities (e.g. mean temperature, measuring room class, etc.). This
option influences machine monitoring, but not measuring room
monitoring.
– All sensors connected to the sensor network are listed in the column
"
Assigned hardware sensors
". Sensors that are connected using
wireless components, must first be coupled to the system as described
in chapter 1.7 of the installation instructions. Each sensor is displayed
as a tile that contains the following information:
– Display: Sensor reachable or not reachable.
Sensors that cannot be reached can be deleted by clicking the top
right minus sign.
– Sensor serial number
– Sensor network address