Operation
37
Operating manual TEMPAR console V9.0
626140-9149-911
– The files with the measured values are created in the enabled folder
according to the "Interval measuring data (ms)” (See chapter "3.2
Operation as “Standard” user group") in the enabled folder. Naming
takes place according to the following template:
“PiWebData_YYYYMMDD-hhmmss_<Seriennummer Konsole/
Messzone>_ABCD.zcsv”.
– The PiWeb Auto Importer reads the measuring data files automatically
according to the set test interval from the specified folder.
Subsequently, these files are deleted or moved to the folder for import
backup.
To export all data from the TEMPAR system, activate “Transfer all data to
PiWeb”. Consequently all data are sequentially exported in large zcsv
files. Dependent on the number of data sets, this can take several hours.
PiWeb REST interface:
For direct communication with the PiWeb server, it must be ensured that
TEMPAR can reach the PiWeb PC/PiWeb server via the network. If a
firewall exists between the PiWeb PC and the TEMPAR system, it must be
ensured that the port that was configured in PiWeb server is open (e.g.
port 8080). To establish the connection, the necessary information must
be entered in the "PiWeb REST interface”. Run the connection status test
if there are connection problems. If a sub-folder is to be created in PiWeb
for the corresponding console / cockpit, this can be set under "Create
subcomponent of measuring room”.
The PiWeb REST interface can be used with a PiWeb servers of
version >= 6.2.
To do so, follow the following steps:
– Activate PiWeb REST interface on the PiWeb server.
– Select Activate PiWeb REST interface.
– Enter the IP address of the PiWeb server.
– Enter the port of the PiWeb server (default: 8080).
– If TEMPAR is successfully connected to the PiWeb server, then after a
short time the PiWeb server version must be displayed in TEMPAR (if
necessary, the TEMPAR page must be reloaded, e.g. by switching
between tabs).
– After a brief period, data are created in the PiWeb database.