52
Maintenance
626140-9149-911
Operating manual TEMPAR console V9.0
4 Maintenance
4.1
Calibration cycle
The temperature sensor calibration cycle must be followed. The
recommended service life for the calibrated temperature sensors is one
year. After a calibration cycle ends, the temperature sensors should be
removed and replaced with newly calibrated ones:
1. Deactivate sensor in Configuration > Measuring Room > Sensors.
2. Unscrew the sensor or the sensor electronics from the sensor network.
3. Place the removed temperature sensors individually in shock-proof
packaging and send them back to the manufacturer for calibration.
4. Screw the newly calibrated sensors into the sensor network.
5. Activate sensor in Configuration > Measuring Room > Sensors.
The removed temperature sensors can be sent back to the manufacturer
for re-calibration. Calibration takes approximately three to six weeks
dependent on the sensor type (plus shipping time). Alternatively the
calibration-exchange service can be used (See chapter "5.1 Scopy of
supply and expansions"). In this way, you receive a newly calibrated
sensor network sent to you before the calibration elapses. All sensors can
be replaced and consequently the measuring process can be
recommenced within a very shot time period. Then the sensors with the
lapsed calibration must be sent back to Carl Zeiss 3D Automation.
Contact:
Carl Zeiss 3D Automation GmbH
Phone:
+49 7361 6336 0
in mailto:
4.2
Maintenance
The system does not require any maintenance.
4.3
Spare parts