SonTek/YSI Inc
RiverSurveyor System Manual (May 1, 2007)
86
6.6.8. Coordinate System
For RiverSurveyor applications, the
Coordinate System
field should always be set to
ENU
(East-
North-Up; Earth coordinates) on the
Advanced Settings
tab of the
User Setup
ADP Configuration|User Setup
). For information on velocity coordinate systems, see §
6.6.9. Bottom Track
For RiverSurveyor applications, the
Bottom Track
field should always be set to
Yes
on the
Ad-
vanced Settings
tab of the
User Setup
ADP Configuration|User Setup
).
For more information on bottom-track, refer to §
6.6.10.
Shallow Water Operation
By default, the
Shallow Water Operation
field should be set to
No
on the
Advanced Settings
tab
of the
User Setup
dialog box (
ADP Configuration|User Setup
). Generally, this field is
enabled for transects or conditions where the water is not deep enough to collect an acceptable
number of cells, or in areas with shallow, wide banks. For more information on shallow-water
operation, refer to §
.
6.6.11.
Recorder
If your system has the internal recorder option installed, we recommend you set the
Recorder
field to
Enabled
to turn on the internal recorder. If your system does not have an internal re-
corder, or if you do not wish to store data to the internal recorder (i.e., you only want to store
data to a drive on your computer), set the
Recorder
field to
Disabled
. The
Recorder
field is on the
Advanced Settings
tab of the
User Setup
;
ADP Configuration|User Setup
).
For more information on the internal recorder, refer to §
.
6.6.12.
Temperature Settings
Water temperature is used in the computation of
the speed of sound and can also be used for other
applications.
•
Temperature Mode
– This setting, on the
Ad-
vanced Settings
tab of the
User Setup
dialog
ADP Configuration|User
Setup
), determines whether the temperature
sensor will be active during data collection.
-
Measured
(default) – Use this option
when you want the built-in temperature
sensor to dynamically measure water
temperature. Each profile will have its
own individually calculated speed of
sound.
-
User
– Use this option if you want to
manually enter a temperature value.
•
Default Water Temp
– This value is used in
the calculation of speed of sound. If
Tem-
perature Mode
is set to
User
, you must
manually enter a water temperature value;
otherwise, this field is disabled.
Figure 6-6. User Setup – Advanced Settings
Содержание SonTek RiverSurveyor
Страница 2: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 ii ...
Страница 4: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 iv ...
Страница 14: ......
Страница 24: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 12 ...
Страница 26: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 14 ...
Страница 71: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 59 Figure 4 7 Exporting data from ViewADP ...
Страница 74: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 62 ...
Страница 82: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 70 ...
Страница 143: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 131 Figure 7 39 Discharge Measurement Summary ...
Страница 148: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 136 ...
Страница 162: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 150 ...
Страница 166: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 154 ...
Страница 168: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 156 ...
Страница 182: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 170 ...
Страница 184: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 172 Figure F 1 FlowPack Discharge Rating Report ...
Страница 194: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 182 ...