5
Category/Item
Comments
1. Ship Track/Current Stick plots
Verify velocity and ship-track reference in menu bar (BT or GPS)
Verify depth reference (BT or ES)
Rewind measurement (F9) and step through individual profiles
2. Velocity Magnitude Contour
Scale graph
Check for bad or missing data (scale contour by profile and distance)
Check bed contour profiles, cut off depth, etc.
3. Signal Amplitude Profile Graph
Rewind measurement (F9) and step through individual profiles
No signal amplitude profiles appear to be intermittently lower than others?
Do signal amplitude profiles indicate similar depths?
Do the signals indicate a similar shape (spike) when reflecting off the riverbed?
4. Tabular Display
Check number of profiles (preferably more than 40)
Check elapsed time (preferably more than 2-3 minutes)
Check % measured discharge (preferably greater than 50%)
Edge discharges realistic and have correct sign? (orientation)
Correct track reference?
Reasonable track distance?
Reasonable DMG?
Number of valid cells great than 2?
Water temperature realistic? (temperature should be very stable)
Average heading realistic?
Average pitch and roll realistic?
Salinity correct? (0 ppt = freshwater; 34.5 = salt/sea water)
Correct ADP transducer depth? (review pressure and depth calibration files)
Correct echo sounder depth? (if applicable)
System voltage between 12 and 24 volts?
5. System Hardware Parameters
Is serial number identical to system used in the field?
Does serial number match the field sheet?
Sensor orientation is down?
Is CPU version current?
6. Setup Parameters
File name prefix similar to field notes?
Is profiling range realistic for measurement section?
Correct averaging interval? (5 seconds is preferable)
Is transducer depth method consistent with field measurement notes?
Is transducer depth consistent with observed field measurements?
7. Velocity Contour Plot
Is flow unidirectional? (no reverse/bi-directional flow observed)
8. Time series/Bar Chart parameter displays
Data is consistent, with no spikes, dropouts or large fluctuations?
Is boat speed relatively steady and non-erratic?
Table E-1. Data Quality Review Sheet
Содержание SonTek RiverSurveyor
Страница 2: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 ii ...
Страница 4: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 iv ...
Страница 14: ......
Страница 24: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 12 ...
Страница 26: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 14 ...
Страница 71: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 59 Figure 4 7 Exporting data from ViewADP ...
Страница 74: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 62 ...
Страница 82: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 70 ...
Страница 143: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 131 Figure 7 39 Discharge Measurement Summary ...
Страница 148: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 136 ...
Страница 162: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 150 ...
Страница 166: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 154 ...
Страница 168: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 156 ...
Страница 182: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 170 ...
Страница 184: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 172 Figure F 1 FlowPack Discharge Rating Report ...
Страница 194: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 182 ...