SonTek/YSI Inc
RiverSurveyor System Manual (May 1, 2007)
10
point, the values displayed are meaningless, especially if your ADP is pinging “in air”
rather than in water. Note: To learn which data is displayed in each column, refer to the
ADP Operation Manual
.
>
start
30212096 free bytes left in recorder.
Free space is sufficient for 8.66 days of operation.
Recorder mode is NORMAL.
Recording data to file DEF030
OK
16 TEST 1 2003 2 13 13 42 51 3 0 0 1
10 400 100 1 9
0 0 0 0 0 0 0 0 0
0 15209
0 0 0 0 118 123 122 127 0 0 0 0 156 246 0 119
1 -7 7 0 7 12 2 0 0 0
2 -7 5 -1 7 5 2 0 0 0
3 8 -3 -3 5 7 1 0 0 0
4 -5 -5 0 10 6 2 0 0 0
5 -9 -9 0 6 6 2 0 0 0
6 10 3 1 8 8 2 0 0 0
7 -1 3 2 11 5 3 0 0 0
8 -7 11 0 5 9 2 0 0 0
9 4 -13 -1 6 4 3 0 0 0
10 9 -5 2 5 5 3 0 0 0
16 TEST 2 2003 2 13 13 42 56 3 0 0 1
10 400 100 1 12
0 0 0 0 0 0 0 0 0
0 15209
0 0 0 0 118 123 122 127 0 0 0 0 156 246 0 119
1 87 -24 63 79 104 19 0 0 0
2 -18 -71 47 138 91 24 0 0 0
3 110 -124 -24 148 87 32 0 0 0
4 -17 217 125 186 209 52 0 0 0
5 59 79 -47 111 53 22 0 0 0
6 42 15 1 88 55 25 0 0 0
7 -12 23 -25 85 83 25 0 0 0
8 -22 -96 -27 160 50 38 0 0 0
9 -9 5 0 94 34 18 0 0 0
10 -40 -89 -4 39 104 43 0 0 0
_
To stop the ADP data collection process, click
Break
(or press
Alt+B
) to send the initializa-
tion (i.e., wake-up) command to the ADP. You should see a copyright/initialization message
similar to the one shown in
<Enter>
.
Note
: The reason we pressed
<Enter>
is to prevent the ADP from resuming the
data collection process. This is because, depending on how your system is config-
ured, if you do not send another command to the ADP within three minutes of
sending the
Break
, the system will once again start collecting data. This is a safety
precaution that helps to resume a deployment in case extraneous signals received
by the ADP are incorrectly interpreted as a
Break
(e.g., when a communication ca-
ble is being connected/disconnected, or power to the system is lost temporarily).
_
Now let’s do a quick check of the compass. Type in
COMPASS CONT
(for “compass con-
tinuous” test) and press
<Enter>
. The
SonUtils
screen will begin to show Heading, Pitch,
and Roll values continuously at a one-second update rate.
Содержание SonTek RiverSurveyor
Страница 2: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 ii ...
Страница 4: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 iv ...
Страница 14: ......
Страница 24: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 12 ...
Страница 26: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 14 ...
Страница 71: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 59 Figure 4 7 Exporting data from ViewADP ...
Страница 74: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 62 ...
Страница 82: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 70 ...
Страница 143: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 131 Figure 7 39 Discharge Measurement Summary ...
Страница 148: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 136 ...
Страница 162: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 150 ...
Страница 166: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 154 ...
Страница 168: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 156 ...
Страница 182: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 170 ...
Страница 184: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 172 Figure F 1 FlowPack Discharge Rating Report ...
Страница 194: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 182 ...