ACTION:
Empty waste bottle.
SECTION:
POSSIBLE CAUSE:
Bubbles in buffer or calibrator bottle tubing.
ACTION:
Prime bottle.
SECTION:
SYMPTOM:
INTERNAL FAILURE: Unable to Calibrate
POSSIBLE CAUSE:
Air bubbles in calibrator tubing.
ACTION:
Prime calibrator bottles.
SECTION:
POSSIBLE CAUSE:
Pinched, leaking or disconnected tube.
ACTION:
Fix or replace tubing.
SECTION:
POSSIBLE CAUSE:
Sipper misaligned.
ACTION:
Check Sipper alignment.
SECTION:
POSSIBLE CAUSE:
Stir bar not in sample module.
ACTION:
Install stir bar.
SECTION:
POSSIBLE CAUSE:
Stir speed too fast or too slow.
ACTION:
Adjust stir speed.
SECTION:
POSSIBLE CAUSE:
Newly installed enzyme membrane.
ACTION:
Enter probe diagnostics and check probe currents.
SECTION:
POSSIBLE CAUSE:
Newly installed probe.
ACTION:
Enter probe diagnostics and check probe currents.
SECTION:
POSSIBLE CAUSE:
Calibrator solution out of spec: contaminated or installed for more than 30 days.
ACTION:
Install new calibrator.
SECTION:
Install Calibrator Solution(s)
POSSIBLE CAUSE:
Net calibration current (PL current) below 5 nA.
ACTION:
Replace enzyme Membrane and check calibrator solution.
SECTION:
POSSIBLE CAUSE:
Failing enzyme membrane.
ACTION:
Perform daily operational checks and replace membrane(s) if necessary.
SECTION:
5.1 Perform Daily Operational Checks
POSSIBLE CAUSE:
Probe surface fouled.
ACTION:
Clean probe surface.
SECTION:
POSSIBLE CAUSE:
Auxiliary electrode fouled.
ACTION:
Clean Auxiliary electrode surface (temperature probe) with isopropyl alcohol.
SYMPTOM:
ERROR: No Fluid Detected
POSSIBLE CAUSE:
Low sample level.
ACTION:
Increase sample level in test tube.
POSSIBLE CAUSE:
Sipper Depth set too high at sample station.
ACTION:
Adjust sipper depth.