124
Sipper pump head
Figure 9.4
Pull the white pump base from the clear pump housing. If salt deposits have formed on the parts, wash them off with
warm water. Make sure the metal pipes where the tubing connects are not blocked or restricted. Replace the seals as
shown in Figure 9.5. Extra seals are supplied in the Preventive Maintenance Kit. Be sure to reinstall the black spacer
between the two small O-rings.
Sipper pump seal replacement
Figure 9.5
Reassemble the pump, position the plunger as shown in Figure 9.6 and install it back on the instrument.
Sipper pump plunger position
Figure 9.6
SPACER