131
Calibrator and Waste Tubing (Modules 2 and 3)
Figure 9.14
9.4.10 Install Sample Modules
After cleaning, flush the sample modules with copious amounts of warm water, then rinse with distilled water to remove
any traces of the disinfecting agent. Install the sample modules. Remember to install the stir bar and the module seal O-
ring. Secure the sample modules using the three thumb nuts. Install new o-rings on the temperature probes, then install
the temperature probes into each sample module.
Be sure to clean the enzyme probes before installing them in the sample module.
9.4.11 Sipper Replacement
Caution: Touch a bare metal chassis screw before handling the sipper tube
to prevent possible damage due to an electrostatic discharge.
The Sipper can be damaged if it is not properly aligned or if its alignment is disturbed. Inspect the Sipper for straightness
and condition of the Teflon jacket separating the two stainless steel tubes. If the Teflon jacket is torn, replacement of the
Sipper is required. Follow the steps below to replace the Sipper.
From the Service screen, Sipper tab, move the sipper to Location [Station 1-P96] to allow access to the sipper.
Remove the screw holding the sipper arm cover, then raise the cover up and slide it out of the sipper arm. Disconnect the
tubing from the sipper.
Screw
Cover
Connect waste
tubing 2 to fittings
Connect waste
tubing 3 to fittings