Connect 2960 Online Monitor
4.4
If you will be using the 2960 Online Monitor, see Section 6.1 for instructions on how to connect it.
Align Sipper
4.5
It is very important that the sipper be accurately adjusted.
WARNING: Keep your hands clear of the sipper while the instrument is in operation.
1. From the Sipper tab of the Service screen, touch [Module 1].
2. The Select Location screen will appear. Select [Module 1]. The sipper will move to sample module 1 and should
be centered above the cone shaped opening in the top of the module.
If the sipper does not move, make sure the packing material was removed.