97
7.4.2.5 Clear Database
NOTES: Make sure all sample data and audit logs have been exported before clearing the database!
The Clear Database function is not available when 21 CFR Part 11 Mode is enabled.
Press [Clear Database] to clear all saved sample and calibration data. Clearing old data from the database will reduce the
amount of time required to export and save sample data.
7.4.2.6 Update Software
From the Help screen, Software tab, insert the flash drive containing the software update in the instruments’ USB port.
After the Flash Drive Inserted light changes to green, touch the [Launch Updater] button.
The 2900 Series Updater will be displayed.
Touch the [Update Firmware] button to install the new firmware.
Touch the [Update Application] button to install the new software.
When the Installation Complete window is displayed, remove flash drive.
The instrument will restart.
NOTE: Updating the software will clear all sample data. After updating the software, the sipper must be aligned
with all positions.
7.4.3 FAQ
From the Help menu, touch the [FAQ] tab to display frequently asked questions and answers.
7.4.4 Training
From the Help menu, touch the [Training] tab.