51
7. Touch the [On] button and change it to [Off] if you are not using and antiseptic solution.
8. Touch [Save] to save your change and return to the Edit Configuration screen.
6.5.4 Filtrate Pump
The filtrate pump control output from the 2960 is provided to control a user-supplied external pump for pumping the
sample from its point of origin to the 2960. The timing of this signal is the same as the purge pump of the 2960.
This
option is normally not used
since the 2960 pump can transport most samples to the Monitor sample cup. If an external
filtrate pump is used, the flow rate must be maintained at or above 580uL/minute in order to guarantee that the sipper
does not aspirate air during the sampling process.
The default filtrate pump output is Active High (+5V). Touch the [Filtrate Pump Active High] button to change it to Active
Low if your pump requires an Active Low signal (0V) to turn on.
6.5.5 Module
Set up a Module for the chemistries you want to monitor/control.
9. Touch the Module 1 [Edit] button.