2.5 Setting the Valid Measurement Range
Setting the Valid Voltage Measurement Range (Voltage)
4.
Tap
Voltage
. A valid voltage measurement range setup screen appears.
Voltage button
Specify the measurement range to switch to when a peak over-range occurs.
(OFF, 1000V, 600V, 300V, 150V, 100V, 60V, 30V, 15V, 10V, 6V, 3V, 1.5V)
• When Range Σ Link is set to on (see section 2.1), the input elements that are assigned to the same
wiring unit are set to the same range.
• If auto range is on (you can turn it on by pressing AUTO), the instrument operates as follows:
• When a peak over-range occurs, the measurement range increases to the range specified here,
skipping the ranges in between.
• When the measurement range to switch to when a peak over-range occurs has not been selected,
the measurement range increases in the order of the specified valid measurement ranges.
Button for
displaying the valid
measurement range
setup screen
Currently specified valid measurement ranges
Measurement ranges that are not valid appear dimmed.
The background of the specified measurement range to switch
to when a peek over-range occurs (Peak Over Jump) is
displayed in blue.