Configuring Motor/Aux Settings (Ch Settings)
The following four types of setup screens are available depending on the motor evaluation
configuration setting.
Single Motor (Speed:Pulse)
4.
Tap
Ch Settings
. A single motor (revolution signal: pulse) setup screen appears.
Ch Settings button
Shows the torque signal input range.
This appears when Sense Type is set to Analog.
Single motor (revolution signal: pulse) setup screen
Set the scaling factor (0.0001 to 99999.9999).
Set the scaling factor that is used to convert the signal from the revolution sensor or
torque meter to speed (rotating speed), torque, and Pm (motor output).
Set the unit (up to 8 characters).
Set the speed, torque, and Pm units.
Set the input signal type (Analog, Pulse).
Set the type of signal (Torque) applied from the torque meter.
Turns auto range on and off
Set the fixed range (20V, 10V, 5V, 2V, 1V).
You can set these when Sense Type is set to Analog.
Set the linear scale
(A: 1.000 m to 1.000 M; B: −1.000 M to 1.000M).
Set A (the slope) and B (the offset).
Computes A and B
Set the line filter (OFF, 1kHz, 500Hz, 100Hz).
9.1 Configuring Motor Evaluation and Auxiliary Input Settings