5-45
IM 760101-11E
Commands
5
5.9
HOLD Group
The commands in this group deal with the hold function of output data.
You can make the same settings and inquiries as when HOLD on the front panel is used.
5.10 IMAGe Group
The commands in this group deal with the output of screen image data.
There are no front panel keys that correspond to the commands in this group.
:IMAGe
:
;
FORMat
<Space>
TIFF
BMP
?
COLor
<Space>
OFF
COLor
REVerse
?
SEND
?
?
:HOLD
<Space>
OFF
ON
<NRf>
?
:HOLD
Function
Sets the output data (display, communications,
etc.) hold or queries the current setting.
Syntax
:HOLD {<Boolean>}
:HOLD?
Example
:HOLD OFF
:HOLD? -> :HOLD 0
:IMAGe?
Function
Queries all settings related to the output of
screen image data.
Syntax
:IMAGe?
Example
:IMAGE? -> :IMAGE:FORMAT TIFF;
COLOR OFF
:IMAGe:COLor
Function
Sets the color tone of the screen image data to
be output or queries the current setting.
Syntax
:IMAGe:COLor {OFF|COLor|REVerse}
:IMAGe:COLor?
Example
:IMAGE:COLOR OFF
:IMAGE:COLOR? -> :IMAGE:COLOR OFF
:IMAGe:FORMat
Function
Sets the output format of the screen image data
or queries the current setting.
Syntax
:IMAGe:FORMat {TIFF|BMP}
:IMAGe:FORMat?
Example
:IMAGE:FORMAT TIFF
:IMAGE:FORMAT? -> :IMAGE:
FORMAT TIFF
:IMAGe:SEND?
Function
Queries the screen image data.
Syntax
:IMAGe:SEND?
Example
:IMAGE:SEND? -> #6
(number of bytes, 6
digits)(series of data bytes)
Description • The number of bytes of <Block data> is {2 + 6
+ number of data 1 (delimiter)}.
• For details on <Block data>, see page 4-6.
Содержание wt1600
Страница 1: ...Digital Power Meter Communication Interface IM 760101 11E 4th Edition ...
Страница 131: ...7 2 IM 760101 11E 7 2 Sample Program Image ...
Страница 138: ...Sample Program 7 9 IM 760101 11E 7 7 4 Output of Normal Measurement Data ...
Страница 142: ...Sample Program 7 13 IM 760101 11E 7 7 5 Output of Harmonic Measurement Data ...