Appendix
App-3
IM 760101-11E
App
Error in communications execution (200 to 299)
Code
Message
Action
Reference Page
221
Setting conflict
Check the relevant setting.
Chapter 5
222
Data out of range
Check the setting range.
Chapter 5
223
Too much data
Check the data byte length.
Chapter 5
224
Illegal parameter value
Check the setting range.
Chapter 5
241
Hardware missing
Check availability of options.
—
260
Expression error
Equation is not allowed.
—
270
Macro error
Does not conform to the macro function specified in IEEE488.2. —
272
Macro execution error
Does not conform to the macro function specified in IEEE488.2. —
273
Illegal macro label
Does not conform to the macro function specified in IEEE488.2. —
275
Macro definition too long
Does not conform to the macro function specified in IEEE488.2. —
276
Macro recursion error
Does not conform to the macro function specified in IEEE488.2. —
277
Macro redefinition not allowed
Does not conform to the macro function specified in IEEE488.2. —
278
Macro header not found
Does not conform to the macro function specified in IEEE488.2. —
Error in communications Query (400 to 499)
Code
Message
Action
Reference Page
410
Query INTERRUPTED
Check transmission/reception order.
4-2
420
Query UNTERMINATED
Check transmission/reception order.
4-2
430
Query DEADLOCKED
Limit the length of the program message including
4-2
<PMT> to 1024 bytes or less.
440
Query UNTERMINATED after
Do not enter any query after
*IDN?
and
*OPT?
.
—
indefinite response
Error in System Operation (912 to 914)
Code
Message
Action
Reference Page
912
Fatal error in Communications-
Servicing is required.
—
driver
Warning
Code
Message
Action
Reference Page
5
*OPC/?
exists in message
Place the
*OPC
or
*OPC?
at the end of the program message.
—
Other errors (350 and 390)
Code
Message
Action
Reference Page
350
Queue overflow
Read the error queue. Code 350 occurs when the error queue
6-5
is full up. This message is output only for the
STATus:ERRor?
query and is not displayed on the screen.
390
Overrun error
Execute with a lower baud rate.
—
(only Serial(RS-232))
Note
Code 350 indicates overflow of error queue. This code is returned as a response to the
“
STATus:ERRor?
” query; it does not appear on the screen.
Appendix 2 Error Messages
Содержание wt1600
Страница 1: ...Digital Power Meter Communication Interface IM 760101 11E 4th Edition ...
Страница 131: ...7 2 IM 760101 11E 7 2 Sample Program Image ...
Страница 138: ...Sample Program 7 9 IM 760101 11E 7 7 4 Output of Normal Measurement Data ...
Страница 142: ...Sample Program 7 13 IM 760101 11E 7 7 5 Output of Harmonic Measurement Data ...