5-73
IM 760101-11E
Commands
5
5.16 RATE Group
The commands in this group deal with the data update rate.
You can make the same settings and inquiries as when UPDATE RATE on the front panel is used.
:RATE
<Space>
<Time>
?
:RATE
Function
Sets the data update rate for normal
measurement or queries the current setting.
Syntax
:RATE {<Time>}
:RATE?
<Time> = 50, 100, 200, 500 (ms), 1, 2, 5 (s)
Example
:RATE 200MS
:RATE? -> :RATE 200.0E-03
Description • This command is valid only during normal
measurement.
• For details on the data update rate during
harmonic measurement, see the WT1600
User’s Manual.
5.17 STATus Group
The commands in the STATus group are used to make settings and inquiries related to the status report.
There are no front panel keys that correspond to the commands in this group.
For details on the status report, see chapter 6.
:STATus
:
;
CONDition
?
EESE
<Space>
<Register>
?
EESR
?
ERRor
?
FILTer
<x>
<Space>
RISE
FALL
BOTH
NEVer
?
QENable
<Space>
OFF
ON
<NRf>
?
QMESsage
<Space>
OFF
ON
<NRf>
?
SPOLl
?
?