5-26
IM 760101-11E
:DISPlay[:NUMeric]:NORMal:IAMount
Function
Sets the numerical display format for normal
measurement or queries the current setting.
Syntax
:DISPlay[:NUMeric]:NORMal:
IAMount {<NRf>|ALL}
:DISPlay[:NUMeric]:NORMal:IAMount?
<NRf> = 4, 8, 16, 42, 78
Example
:DISPLAY:NUMERIC:NORMAL:IAMOUNT 4
:DISPLAY:NUMERIC:NORMAL:IAMOUNT? ->
:DISPLAY:NUMERIC:NORMAL:IAMOUNT 4
Description The contents of the measured data that are
displayed are as follows depending on the
setting of the numerical display format.
<NRf>: Numerical display items are displayed
in order by the item number.(<NRf> expresses
the number of items that is displayed on a single
screen.)
ALL: All functions are displayed in order by
element.
:DISPlay[:NUMeric]:NORMal:ICURsor
Function
Sets the cursor position on the numerical
display (split display) for normal measurement
or queries the current setting.
Syntax
:DISPlay[:NUMeric]:NORMal:
ICURsor {<NRf>}
:DISPlay[:NUMeric]:NORMal:ICURsor?
<NRf> = 1 to 100
Example
:DISPLAY:NUMERIC:NORMAL:ICURSOR 1
:DISPLAY:NUMERIC:NORMAL:ICURSOR? ->
:DISPLAY:NUMERIC:NORMAL:ICURSOR 1
Description • Specify the cursor position in terms of the
item number.
• This command is valid when the display
format of numeric values (
:DISPlay
[:NUMeric]:NORMal:IAMount
) is set to
“<NRf> (split display).”
:DISPlay[:NUMeric]:NORMal:ITEM<x>
Function
Sets the numerical display item for normal
measurement or queries the current setting.
Syntax
:DISPlay[:NUMeric]:NORMal:ITEM<x>
{NONE|<Function>,<Element>}
:DISPlay[:NUMeric]:NORMal:ITEM<x>?
<x> = 1 to 100 (item number)
NONE = No display item
<Function> =
{URMS|UMN|UDC|UAC|IRMS|
...}
(See the function selection list (1).”)
<Element> =
{<NRf>|SIGMA|SIGMB|SIGMC}
(<NRf> = 1 to 6)
Example
:DISPLAY:NUMERIC:NORMAL:
ITEM1 URMS,1
:DISPLAY:NUMERIC:NORMAL:ITEM1? ->
:DISPLAY:NUMERIC:NORMAL:
ITEM1 URMS,1
Description This command is valid when the display format
of numeric values
(:DISPlay[:NUMeric]
:NORMal:IAMount
) is set to “<NRf> (split
display).”
:DISPlay[:NUMeric]:NORMal:PRESet
Function
Presets the display order pattern of numerical
display items for harmonic measurement.
Syntax
:DISPlay[:NUMeric]:NORMal:
PRESet {<NRf>}
<NRf> = 1 to 4
Example
:DISPLAY:NUMERIC:NORMAL:PRESET 1
Description Regardless of what value (1 to 4) is specified for
<NRf>, the display pattern (order) of the
numerical display items will be the same as the
display order when Reset List Exec of the
Display setting menu, which is displayed on the
WT1600 screen, is executed.For details on the
order of displayed items when reset is executed,
see the WT1600 User’s Manual.
:DISPlay:TRENd?
Function
Queries all settings related to the trend.
Syntax
:DISPlay:TRENd?
Example
:DISPLAY:TREND? -> :DISPLAY:TREND:
SAMPLING 1;T1 1;T2 1;T3 1;T4 1;
T5 1;T6 1;T7 1;T8 1;T9 0;T10 0;
T11 0;T12 0;T13 0;T14 0;T15 0;
T16 0;FORMAT SINGLE;TDIV 0,0,3;
NORMAL:ITEM1:FUNCTION URMS,1;
SCALING:MODE AUTO;VALUE 100.00E+00,
-100.00E+00;:DISPLAY:TREND:NORMAL:
ITEM2:FUNCTION IRMS,1;SCALING:
MODE AUTO;VALUE 100.00E+00,
-100.00E+00;...(omitted)...;:
DISPLAY:TREND:NORMAL:ITEM16:
FUNCTION FU,2;SCALING:MODE AUTO;
VALUE 100.00E+00,-100.00E+00
5.5 DISPlay Group
Содержание wt1600
Страница 1: ...Digital Power Meter Communication Interface IM 760101 11E 4th Edition ...
Страница 131: ...7 2 IM 760101 11E 7 2 Sample Program Image ...
Страница 138: ...Sample Program 7 9 IM 760101 11E 7 7 4 Output of Normal Measurement Data ...
Страница 142: ...Sample Program 7 13 IM 760101 11E 7 7 5 Output of Harmonic Measurement Data ...