5-64
IM 760101-11E
:MOTor:SPEed:PULSe
Function
Sets the pulse count of the revolution sensor
signal input (pulse input) or queries the current
setting.
Syntax
:MOTor:SPEed:PULSe {<NRf>}
:MOTor:SPEed:PULSe?
<NRf> = 1 to 9999
Example
:MOTOR:SPEED:PULSE 60
:MOTOR:SPEED:PULSE? -> :MOTOR:
SPEED:PULSE 60
Description This command is valid when the input type of
the revolution sensor signal (
:MOTor:SPEed:
TYPE
) is set to “PULSe (pulse input).”
:MOTor:SPEed:RANGe
Function
Sets the voltage range of the revolution sensor
signal input or queries the current setting.
Syntax
:MOTor:SPEed:RANGe {<Voltage>|AUTO}
:MOTor:SPEed:RANGe?
<voltage> = 1, 2, 5, 10, and 20 (V)
AUTO = Auto range
Example
:MOTOR:SPEED:RANGE 20V
:MOTOR:SPEED:RANGE? -> :MOTOR:
SPEED:RANGE 20.0E+00
Description You cannot specify this command when the
input type of the revolution sensor signal
(
:MOTor:SPEed:TYPE
) is set to “PULSe (pulse
input).” The amplitude input range of the pulse
signal is
±
5 Vpeak.
:MOTor:SPEed:SCALing
Function
Sets the scaling factor for rotating speed
computation or queries the current setting.
Syntax
:MOTor:SPEed:SCALing {<NRf>}
:MOTor:SPEed:SCALing?
<NRf> = 0.0001 to 99999.9999
Example
:MOTOR:SPEED:SCALING 1
:MOTOR:SPEED:SCALING? -> :MOTOR:
SPEED:SCALING 1.0000
:MOTor:SPEed:TYPE
Function
Sets the input type of the revolution sensor
signal input or queries the current setting.
Syntax
:MOTor:SPEed:TYPE {ANALog|PULSe}
:MOTor:SPEed:TYPE?
Example
:MOTOR:SPEED:TYPE ANALOG
:MOTOR:SPEED:TYPE? -> :MOTOR:SPEED:
TYPE ANALOG
:MOTor:SPEed:UNIT
Function
Sets the unit to add to the rotating speed
computation result or queries the current
setting.
Syntax
:MOTor:SPEed:UNIT {<String>}
:MOTor:SPEed:UNIT?
<string> = 8 characters or less
Example
:MOTOR:SPEED:UNIT "rpm"
:MOTOR:SPEED:UNIT? -> :MOTOR:SPEED:
UNIT "rpm"
Description • Only the characters and symbols displayed
on the keyboard on the screen can be used.
• This command does not affect the
computation result.
:MOTor:SSPeed (Sync SPeed source)
Function
Sets the frequency measurement source used
to compute the synchronous speed (SyncSpd)
or queries the current setting.
Syntax
:MOTor:SSPeed {U<x>|I<x>}
:MOTor:SSPeed?
<x> = 1 to 6 (element)
Example
:MOTOR:SSPEED I1
:MOTOR:SSPEED? -> :MOTOR:SSPEED I1
:MOTor:SYNChronize
Function
Sets the synchronization source used to
compute the rotating speed and torque or
queries the current setting.
Syntax
:MOTor:SYNChronize {U<x>|I<x>|
EXTernal|NONE}
:MOTor:SYNChronize?
<x> = 1 to 6 (element)
EXTernal = External clock input (Ext Clk)
NONE = No synchronization source
Example
:MOTOR:SYNCHRONIZE NONE
:MOTOR:SYNCHRONIZE? -> :MOTOR:
SYNCHRONIZE NONE
:MOTor:TORQue?
Function
Queries all settings related to the torque meter
signal input.
Syntax
:MOTor:TORQue?
Example
:MOTOR:TORQUE? -> :MOTOR:TORQUE:
RANGE 20.0E+00;SCALING 1.0000;
UNIT "Nm"
:MOTor:TORQue:RANGe
Function
Sets the voltage range of the torque meter
signal input or queries the current setting.
Syntax
:MOTor:TORQue:RANGe {<Voltage>|AUTO}
:MOTor:TORQue:RANGe?
<voltage> = 1, 2, 5, 10, 20, and 20 (V)
AUTO = Auto range
Example
:MOTOR:TORQUE:RANGE 20V
:MOTOR:TORQUE:RANGE? -> :MOTOR:
TORQUE:RANGE 20.0E+00
5.14 MOTor Group
Содержание wt1600
Страница 1: ...Digital Power Meter Communication Interface IM 760101 11E 4th Edition ...
Страница 131: ...7 2 IM 760101 11E 7 2 Sample Program Image ...
Страница 138: ...Sample Program 7 9 IM 760101 11E 7 7 4 Output of Normal Measurement Data ...
Страница 142: ...Sample Program 7 13 IM 760101 11E 7 7 5 Output of Harmonic Measurement Data ...