<6. SETTING AND CALIBRATION>
6-31
IM 11G02N01-01E
Meaning of setting
Description of setting
ENT
( )
ENT
( )
ENT
End of Hold Setting
Parameter Setting screen
ESC
F6_7_6.ai
5) On the parameter hold screen that
appears, move the cursor next to the Ch
(component) you want to make the
setting by pressing the
or the
key,
and then press the
ENT
key.
6) The value is highlighted, indicating that
the value can be changed. Change the
value by pressing the
or the
key,
and then move the cursor to the right by
pressing the
key.
7) After the value is changed, press the
ENT
key.
The setting is expressed in % against the range
for both ranges.
When 0 to 1000 ppm is selected as the range,
for example, if 10% FS is selected as hold
setting, the output equivalent to 100 ppm is
output and held irrespective of the measurement
value at that time.
8) Press the
ESC
key to return to the
parameter setting screen.
• Instantaneous value display of the measurement cannot be held. (Output only can be held.)
• If set value is selected for hold, instantaneous O
2
correction value is calculated and held
based on the set value.
• Range identification contact output cannot be switched even if the range is switched during
the hold.
Содержание IR400 series
Страница 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Страница 7: ...Blank Page ...
Страница 11: ...Blank Page ...
Страница 33: ...Blank Page ...
Страница 35: ...Blank Page ...
Страница 83: ...Blank Page ...
Страница 93: ...Blank Page ...
Страница 97: ...Blank Page ...
Страница 117: ...Blank Page ...