<6. SETTING AND CALIBRATION>
6-7
IM 11G02N01-01E
6.2.3 Setting of calibration range
This mode is used to set if the range of each Ch (component) at the calibration (manual calibration or
auto calibration) should be calibrated with a single range or 2 ranges.
MODE
ENT
( )
ENT
( )
ENT
( )
F6_2_3.ai
(1) During measurement, press the
MODE
key
to display the User Mode.
(2) Point the cursor to “Calibration
Parameters” by pressing the
or
key. Press the
ENT
key.
(3) In the “Calibration Parameters” screen
that appears, point the cursor to “About
Calibration Range” by pressing the
or
key. Press the
ENT
key.
(4) In the “Calibration Range Ch Selection”
screen that appears, point the cursor to the
Ch you want to set by pressing the
or
key. Press the
ENT
key.