<6. SETTING AND CALIBRATION>
6-15
IM 11G02N01-01E
MODE
ENT
( )
ENT
End of Gas Flow Time Setting
F6_4_1_2.ai
<Gas flow time> setting
(1) Press the
ENT
key in a state where the
cursor is placed next to “Flow Time,” and
the flow time setting screen shown at right
appears.
(2) On the flow time setting screen that
appears,
move the cursor to the gas you want to change
the setting by pressing the
or
the key,
and then press the
ENT
key.
(3) The highlighted value can be changed.
Change the value by pressing the
or
key, and then move the cursor to the right
by pressing the
key.
(4) After changing the value, press the
ENT
key.
(5) Press the
ESC
key to return to the automatic
calibration setting screen.
Note) Only the Chs used are displayed on this screen.
The Ex. time is the output signal hold extension time after
the completion of calibration.
It is valid only when the hold setting is set to “ON.”
The Ex. time set here is also the hold extension time at the
time of manual calibration.
Содержание IR400 series
Страница 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Страница 7: ...Blank Page ...
Страница 11: ...Blank Page ...
Страница 33: ...Blank Page ...
Страница 35: ...Blank Page ...
Страница 83: ...Blank Page ...
Страница 93: ...Blank Page ...
Страница 97: ...Blank Page ...
Страница 117: ...Blank Page ...