<9.SPECIFICATIONS>
9-15
IM 11G02N01-01E
STANDARD ACCESSORIES
Name
Pert Number
Description
Qty
Power cable
K9218SA
standard inlet type (2.5 m )
1
Fuse
K9218SB
replacement fuse (250 V AC, 3.15 A, delay type) ×1
2
Input/Output
terminal module
K9218SC
External terminal module
1
Cable
K9218SD
Connection cable between main unit
and input/output terminal module (1 m)
1
Slide rail
K9218SE
Slide rail
2
Note: Quantity in this is number of accessories supplied as standard. For instance, two K9218SE parts, i.e.,
two slide rails, are supplied as standard. When ordering separately, the required number of should be considered.
Dedicated Zirconia O
2
Sensor (to be purchased separately)
For O
2
correction, the IR400 can accept linearized 0 to 1 V DC signal coming from an analyzer
calibrated to 0 to 25%O
2
of full scale. Dedicated zirconia O
2
sensor , Model ZX8D, is available from
Yokogawa.
Measuring method: Zirconia system Measurable component and measuring range:
Measurable component
Minimum range
Maximum range
Oxygen (O
2
)
0-5 vol%
0-25 vol%
Repeatability:
Within ± 0.5% of full scale
Linearity:
Within ± 1% of full scale
Zero drift:
Within ± 1% of full scale/week
Span drift:
Within ± 2% of full scale/week
Response time:
Approx. 20 seconds (for 90% response)
Measured gas flow rate: 0.5 ± 0.25 L / min
Remark:
The Zirconia system, due to its principle, may produce a measuring error
due to relative concentration versus the combustible O
2
gas concentration.
Also, a corrosive gas (SO
2
of 250 ppm or more, etc.) may affect the life of
the
sensor.
Gas inlet/outlet size:
Rc1/4
Power supply:
90 to 126 V AC or 200 to 240 V AC, 50/60 Hz
Enclosure:
Steel casing, for indoor application
Indication:
Temperature indication (LED)
Temperature alarm output:
Contact output
1a contact,
Contact capacity
220 V AC, 1 A (resistive load)
Safety and EMC conforming standards:
Installation altitude:
2000 m or less
Pollution degree :
2 (Note)
Installation category:
II (Note)
Note:
Installation category, called over-voltage category, specifies impulse withstanding voltage.
Category II is for electrical equipment. Pollution degree indicates the degree of existence of solid,
liquid, gas or other inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.
Safety:
EN61010-1
EMC:
EN61326-1 Class A, (For use in industrial locations),
EN61326-2-3, EN61000-3-2, EN61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand
Note:
The product mounted in a steel enclosure conforms to the requirements of EMC directive.
Содержание IR400 series
Страница 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Страница 7: ...Blank Page ...
Страница 11: ...Blank Page ...
Страница 33: ...Blank Page ...
Страница 35: ...Blank Page ...
Страница 83: ...Blank Page ...
Страница 93: ...Blank Page ...
Страница 97: ...Blank Page ...
Страница 117: ...Blank Page ...