<6. SETTING AND CALIBRATION>
6-22
IM 11G02N01-01E
6.5.2 Forced run/stop of auto zero calibration
Auto zero calibration can be performed just once, or auto zero calibration can be forcibly stopped
during calibration.
6.5.2.1 Execution of auto zero calibration (just once)
ENT
( )
ENT
( )
F6_5_2.ai
(1) Move the cursor to “Setting of Auto Zero
Calibration” by pressing the
or
key on the User Mode screen, and
then press the
ENT
key.
(2) In the “Setting of Auto Zero Calibration”
item selection screen that appears, point
the cursor to “Auto Zero Calibration
Run” by pressing the
or
key.
Press the
ENT
key.
(3) “Run” is highlighted, displaying a message
to confirm execution of auto zero calibration.
Press the
ENT
key to execute the calibration
forcibly, and press the
ESC
key to cancel.