<6. SETTING AND CALIBRATION>
6-17
IM 11G02N01-01E
6.4.2 Forced run/stop of auto calibration
Auto calibration can be performed just once or forcibly stopped while the calibration is performed.
6.4.2.1 Execution of auto calibration (only once)
ENT
( )
ENT
( )
F6_4_2.ai
(1) Display the User Mode screen. Move the
cursor to “Setting of Auto Calibration”
by pressing the
or the
key, and
then press the
ENT
key.
(2) In the “Setting of Auto Calibration” item
selection screen that appears, point the
cursor to “Auto Calibration Run” by
pressing the
or
key.
Press the
ENT
key.
(3) “Run” is highlighted, displaying a message
to confirm the execution of auto calibration.
Press the
ENT
key to execute the auto
calibration forcibly, and press the
ESC
key
to cancel.
Содержание IR400 series
Страница 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Страница 7: ...Blank Page ...
Страница 11: ...Blank Page ...
Страница 33: ...Blank Page ...
Страница 35: ...Blank Page ...
Страница 83: ...Blank Page ...
Страница 93: ...Blank Page ...
Страница 97: ...Blank Page ...
Страница 117: ...Blank Page ...